Find a Distributor
MAC155 Humidity Probe & Humidity Transmitter For Continuous Stack Monitoring With Built in Two Point Calibration
The MAC155 moisture/humidity analyzer enables you to augment your process safety and comply with US EPA regulations. Several high-quality components such as a humidity probe and humidity transmitter combine to make this device exceptionally user-friendly. MAC155 is used in industrial operations worldwide for 24/7 monitoring of stack emission gases. It is designed for measuring humidity in exhaust gases in temperatures ranges up to 1200°F (650°C). MAC Instruments also provides a range of special accessories to expand the usability of the humidity probe in temperatures up to 2400°F (1300°C).
A defining attribute of the MAC155 humidity transmitter is that it features an in-built calibration system. This makes daily 2-point in-situ calibration and system checks possible, and facilitates adherence to US EPA regulations and local mandates pertaining to exhaust stack installations. The humidity probe is protected from particles by a 10 micron stainless steel sintered filter element that can be easily cleaned and replaced.
The humidity probe on this unit is exceedingly sensitive to water vapor/humidity. However, it does not rely on a wet bulb, mirrors, optics, chemicals, or compressed air to provide readings. This humidity meter is highly responsive to the partial pressure of water vapor, thus displaying instantaneous readings. You can confidently depend on MAC Instrument’s patented solid state design to help you improve overall process efficiencies in the toughest industrial settings. The MAC155 humidity transmitter is available in two models for low range and high range continuous humidity monitoring.
- MAC155 Humidity Analyzers is Available in Two Models:
- MAC155-L covers 3% to 15% or 6% to 30% ranges (user selectable)
- MAC155-H covers 3% to 15%, 6% to 30%, 9% to 45%, or 12 % to 60% ranges (user selectable)
- Features of the MAC-155 Moisture/Humidity Transmitter
- Measure % Moisture By Volume In Exhaust Gasses At Temperatures up to 1200°F
- Optional to 2400°F
- NEMA 4X Waterproof Design
- Suitable for outdoor installation
- Contains Built In Equipment Allowing Daily 2 Point Calibration Check
- Sensor is flooded with an internally generated reference gas of specific humidity levels
- Can Be Re-Calibrated In the Field Using Internal Calibrator
- System Calibration
- Sensor signal substitution can be externally initiated to generate an output signal 50% of full output.
- From the MAC155 Controller, 3 different output signals can be generated.
- Provides a method of checking total system including data acquisition equipment.
- Patented Measurement Process
- In Situ Measurement
- Mounts directly to stacks or ducts
- 0-20mA or 4-20mA Signal Output User selectable Self powered or loop powered (jumper selectable)
- Filter Blow Back Option
- Allows automatic filter cleaning (requires 10 psi clean dry air)
- Case Heater Option
- Extend ambient temperature rating to -40°F